SX10507 Demo: Calibration Summary
Before you start:
- Select via the menu mode, the temperature source you intend to use for calibration
- Select via Menu mode, a suitable filter setting to get a stable pH reading in the status screen
- Select via the Menu mode, the temperature units you wish to work in (Fahrenheit or Celsius)
Calibrating:
- Enter calibration mode with long press, Input 8
- pH Probe in PH7 buffer solution, adjust USER temperature (if temperature source is set to USER, Input 12 and Input 11)
- Enter the tabulated buffer pH value for the buffer solution temperature (Input 10 and Input 9)
- Wait for the Reading to be stable
- Press Input 8 button to advance to the pH4 calibration
- pH Probe in PH4 buffer solution, adjust USER temperature (if temperature source is set to USER, Input 12 and Input 11)
- Enter the tabulated buffer pH value for the buffer solution temperature (Input 10 and Input 9)
- Wait for the Reading to be stable
- Press Input 8 button to finish calibration.
Notes:
- Rinse probe in distilled water between dipping it into pH buffers to avoid contamination
- Shake off distilled water from probe to avoid diluting the pH buffer solutions
- Always store the probe in the solution supplied with the probe (KCl solution), NOT distilled water.